This report documents the execution, evaluation, and results of the initial pilot of the FAITH Large Scale Pilot 6 (LSP6) — the Healthcare Large-Scale Pilot — conducted in the context of Task 9.2 (T9.2: Initial Pilot, M3–M26). Building upon the coordination and experimental infrastructure established in D9.1, this deliverable presents the outcomes of the prospective clinical deployment of the AI System Under Test (AI SUT) at the University Hospital of Heraklion, Greece (PAGNH), under the oversight of FORTH, and with contributions from PAGNH, UNIPI, and CNR.
Purpose and Scope of the Initial Pilot. The primary objective of the initial pilot is the real-world clinical evaluation of a deep learning-based prostate gland segmentation model deployed within the radiology workflow of PAGNH. The AI SUT is designed to replace the second reader of multiparametric MRI (mpMRI) images, targeting the delineation of the whole prostate gland, peripheral zone, and transition zone from T2-weighted MRI scans. In doing so, the pilot addresses the persistent clinical challenges of intra- and inter-observer variability in prostate segmentation, seeking to improve diagnostic reproducibility, reduce radiologist workload, and contribute to cost-efficient cancer imaging workflows.
AI System Under Test — Performance and Advancement. Entering the initial pilot at Technology Readiness Level (TRL) 5 with a whole-gland Dice Similarity Coefficient (DSC) of 0.93, the AI SUT has been further refined prospectively on cases acquired at PAGNH during the pilot period. Zonal segmentation performance stands at DSC 0.89 for the transition zone and DSC 0.83 for the peripheral zone, utilising a nnU-Net backbone architecture. Prospective model retraining on pilot collected mpMRI cases — drawn from approximately 9.4 patient cases per month as per the project’s target throughput — has incrementally strengthened generalisation across MRI vendors and field-strength variability (1.5 T and 3 T), directly targeting KPI2. The model is on track for advancement to TRL 6 upon completion of this phase.
Trustworthiness Assessment via the FAITH AI_TAF. The initial pilot operationalises the FAITH AI Trustworthiness Assessment Framework (FAITH AI_TAF) for the first time within a live clinical environment. Trustworthiness assessment activities have been structured across seven dimensions — safety, security and resilience, bias mitigation, explainability and interpretability, accountability and transparency, privacy enhancement, and validity and reliability — in direct alignment with the NIST AI Risk Management Framework and EU AI Act requirements applicable to Software as a Medical Device (SaMD). The completion of this trustworthiness evaluation cycle, generating a documented FAITH AI_TAF risk management profile for the healthcare domain, constitutes KPI3 of LSP6, with its outputs directly informing the replication pilot at UNIPI.
Clinical Validation and KPI Attainment. Evidence gathered through the structured clinical
evaluation protocol at PAGNH — combining quantitative segmentation metrics, expert radiologist feedback via the embedded trustworthiness questionnaire, and qualitative usability assessments — is reported herein. Results indicate that more than 90% of AI-generated segmentations successfully exclude extra-prostatic tissues in line with KPI1. Stakeholder feedback has confirmed the clinical utility and interpretability of the confidence heatmap visualisations integrated into the OHIF DICOM viewer. Regulatory and ethical compliance remains in force throughout, with the Greek Institutional Review Board (IRB) approval underpinning all data processing activities in accordance with GDPR and the EU Medical Device Regulation (MDR 2017/745).